探索全新的 mydSPACE 服务: 许可证管理、全新的 dSPACE 帮助以及用于用户文档的 dSPACE 聊天机器人。请访问我们的专用支持页面,获取更多信息和支持。 链接了解更多并获取支持

FAQ 791: Processing Unit Not Accessible

Question

What to do when the SCALEXIO Processing Unit is not accessible?

Solution

Follow the steps below to identify and resolve the issue.

1. Quick Check – Is the Processing Unit reachable via the web interface?

Open a web browser and enter the IP address of the Processing Unit (PU).

If the PU is accessible:

  • Open the SUPPORT page
  • Generate a Limited Report (*.tgz)
  • Contact dSPACE Support and request an upload link (see step 8)
  • Upload the report for analysis

This is usually the fastest way to resolve the issue.

 

If the PU is not accessible:

Continue with the steps below. They describe which information should be collected when contacting dSPACE Support.

2. Identify Your Hardware

Provide the following information when contacting dSPACE Support:

  • PU type
  • If unknown: serial number (SN) or label information

The label can be found:

  • On the rear side
  • On the left side
  • Behind the front flap (depending on the PU type)

3. Observe Boot Behavior

Check how the system behaves during startup:

  • Record a video of the LEDs (PWR/SYS/APP)
  • Include any acoustic signals (beeps) if possible
  • Alternatively: Compare the LED behavior with Table 1

This information is helpful for troubleshooting.

4. Connect a Monitor

Sending a video of the boot information displayed on the monitor can help dSPACE Support diagnose the issue.Connect a monitor to the PU and restart the system.
Check for:

  • Displayed IP address
  • Error messages
  • Boot messages

The system has booted successfully if you see:

  • An IP address
  • The string “DS-86_32” with firmware information

In this case, continue with network troubleshooting (Step 5).

5. Check the Network Connection

  • Verify the Ethernet cable connection
  • Replace the cable if necessary
  • Ensure the cable is connected to the host interface
  • Ensure the cable is not connected to the 2nd Ethernet port
  • Check for LED activity on the network connector
  • Test with a direct (peer-to-peer) connection
  • Ensure:
    • The host and PU are in the same subnet
    • The PU responds to ping

6. Perform Hardware Checks

If the PU does not boot properly (see step 4, Monitor), check the following components:

6.1 Boot Medium (SSD/CompactFlash)
  • Identify the boot medium used, see Table 2
  • If available:
    → Cross-swap the boot medium with a working system

Additional steps for CompactFlash only:

  • dSPACE Support can provide an image file
  • Rewriting the card often resolves issues
  • If not, the card might be defective 
6.2 DS2502 Router Board
  • The DS2502 IOCNET Board can cause boot issues
  • If available:
    → Test with a replacement board 
6.3 RAM
  • Remove and reinsert the RAM modules
  • Clean the contacts if necessary
  • Insert the bar in another slot only for testing
  • If multiple modules are installed:
    • Test them one by one
    • Remove one to check whether the behavior changes

⚠️ Follow ESD precautions 

6.4 CMOS Battery

Typical symptom: repeated restarts

Procedure:

  1. Disconnect the power supply
  2. Wait at least one minute
  3. Measure the battery voltage (with the battery still installed) at positive pole against housing ground
6.5 Power Supply/Mainboard

If possible:

  • Remove the PU from the HIL system
  • Connect it directly to the mains power supply

If:

  • No LEDs light up
  • The system shows no activity

→ Likely causes:

  • Defective power supply
  • Defective mainboard

 

7. Factory Mode

You can force the PU into factory mode:

  • Connect a keyboard (USB)
  • Press the B key during startup
  • Hold until the SYS LED turns red (~30–40 s)

If it does not work (e.g., on an HCP-P02)

  • Connect also a monitor (Display Port)
  • Watch the screen output
  • Press the B key when you see output similar to that shown in Photo 1

Result:

  • System should become accessible again (as long as it is not defective)
  • A firmware update is required

Note: This does NOT reset the IP nor the configuration

 

8. Contact dSPACE Support

If the issue persists:

  • Contact dSPACE Support
  • If necessary, provide:
  • PU type or serial number
  • LED behavior
  • Photos or videos
  • Report file (if available)

To send the device in for repair, see Hardware Repairs / RMA

Tables

Table 1: Table of LED states from PU:

PWR SYS APP Description

Off

Off

Off

The SCALEXIO system is switched off.

Green

Any

Any

The power supply is OK.

Green

Off (permanent)

Off

If the SCALEXIO system is switched on and the SYS LED is off for longer than 60 s, the SCALEXIO processing hardware is defective. Contact dSPACE Support.

Green

Off (temporary)

Off

The BIOS initialization is running.

Green

Red (permanent)

Off

Error state, contact dSPACE Support.

Green

Red (temporary)

Off

The BIOS initialization is complete. The SCALEXIO RTLib is booting.

Green

Red/orange flashing

Red/orange flashing

The FPGA is loaded. The installed DS2502 IOCNET Link Board is not detected.

Error state, contact dSPACE Support.

Green

Red/orange flashing

Off

The installed DS2502 IOCNET Link Board is detected. The BIOS initialization is still running. SCALEXIO RTLib has not yet booted completely.

If this state is displayed longer than 60 s, the SCALEXIO system cannot boot due to a general boot problem or a defect. Contact dSPACE Support.

Green

Green flashing

Off

SCALEXIO RTLib has booted. The Ethernet connection to the host PC has not been established (yet).

Green

Orange/green flashing

Off

SCALEXIO RTLib is ready. The Ethernet connection to the host PC has been established. The IP address has not been set (yet). The processing unit is waiting for the DHCP server.

Green

Green

Off

The SCALEXIO system is ready for operation. SCALEXIO RTLib is ready. The Ethernet connection to the host PC has been established. The IP address has been set.

No application is loaded.

Green

Green flashing (three times)

Green flashing (three times)

Status message when specifying the network configuration: The network configuration was successful. Shut down and restart the SCALEXIO system to activate the settings.

Green

Red flashing (three times)

Red flashing (three times)

Status message when specifying the network configuration: The SCALEXIO system detected the configuration service, but the configuration was not successful. Ensure that the SCALEXIO system is not accessed by software, such as ConfigurationDesk or ControlDesk, and no real-time application is loaded to the system.

Green

Green

Green

An application is loaded and running.

Green

Green

Orange

An application is loaded. A simulation is running and a connection to the electrical error simulation service is established.

Green

Green

Red

An application loaded is stopped or terminated.

Table 2: Boot medium

PU Variant is Booting From

SSD

CompactFlash (CF) card

HCP P01

 

Yes

HCP P02

 

Yes

HCP P03

 

Yes

HCP P04

Yes

 

HPP P01

 

Yes

HPP P02

Yes

 

HPP P03

Yes

 

ES P01

Yes

 

 

 

 

Photo 1: Monitor output

Related Topics

  • FAQ 361
    FAQ 361

    I do not know the version of the Real-Time PC in my SCALEXIO system. How can I identify the version of the Real-Time PC?

Tags

Date 2026-09-25
产品 SCALEXIO Processing Unit
资料信息 常见问答
信息类别 故障排除

通过 dSPACE时事通讯服务了解最新信息。

通过 dSPACE 时事通讯服务,我们将随时向您通报当前的使用案例、新解决方案和产品以及培训和活动。在此注册,免费订阅。

Enable form call

At this point, an input form from Click Dimensions is integrated. This enables us to process your newsletter subscription. The form is currently hidden due to your privacy settings for our website.

External input form

By activating the input form, you consent to personal data being transmitted to Click Dimensions within the EU, in the USA, Canada or Australia. More on this in our privacy policy.